Online Residual Chlorine Meter
The Online Residual Chlorine Meter is designed for continuous residual chlorine monitoring in water treatment, drinking water distribution, swimming pools and cooling circulating water systems. This residual chlorine analyzer provides stable chlorine measurement for online water quality monitoring and helps users improve disinfection control, dosing accuracy and water safety management.
- Product Introduction
Measurement principle
The CLB-H4130 online residual chlorine meter uses a single microprocessor to measure active chlorine, pH, temperature and residual chlorine conversion.
Technical parameter
|
Measuring range |
Residual chlorine: 0~10.00mg/L (maximum 20mg/L), resolution: 0.001mg/L; Temperature: 0~45℃, resolution: 0.1℃; PH: 0~14PH (if the PH value of water quality is stable, it can be left out when measuring with PH electrode). |
|
Stability |
At room temperature, the drift is less than 2%F S per month. |
|
Isolated output |
0~10mA or 4~20mA |
|
Electronic unit output current error |
±0.1mA |
|
Two sets of relay normally open contacts |
3A/240VAC,6A/28VDC.120VAC. |
|
Electronic unit alarm error |
±0.1%F·S |
|
Power Supply Voltage |
AC: 200~265V/50~60Hz; DC: 24V (DC order must be declared). |
|
Power dissipation |
≤3W |
|
Sensor (residual chlorine electrode) life |
≥ 1 year |
|
Response time |
<2min (90%,20℃) |
|
Polarization time |
≥1h |
|
Minimum flow |
15cm3/S (constant flow rate) |
|
Water sample pressure |
≤0.6MPa |
|
Quality of electronic unit (base type) |
About 600g |
|
Shape and size of electronic unit base |
96 (long) x 96 (wide) x 125 (deep) mm |
|
Opening size for base installation of electronic unit |
91x91mm; the inlet and outlet pipes of the circulation pool are equipped with valves and short pipe connectors with outer diameters Φ8~Φ6, which are connected by flexible hoses; if the electrode is installed in a submerged manner: connecting rods and fixing brackets are purchased, which can be submerged in water for 1~2 meters. |
|
Working conditions |
Environmental temperature: 0~45℃; relative humidity: not more than 90%. |
Functional description
- Intelligent: a single chip microprocessor is used to complete the measurement of residual chlorine value, pH value and temperature;
- Human-computer dialogue interface: menu operation structure, users can operate according to the prompts on the screen;
- Network: can be transmitted, monitored and controlled remotely;
- Isolated output current: software selects 0~10mA or 4~20mA output;
- Measurement range: the upper limit and lower limit parameters correspond to the output current can be set freely;
- Upper and lower limit alarm can be set arbitrarily, and the limit will flash (measurement status);
- Two sets of relays control, with adjustable hysteresis, allowing external connection to metering pumps, solenoid valves, and other control devices for automatic dosing and other controls;Set your own password, users can set or change their passwords to prevent unauthorized personnel from entering and causing incorrect operations; it also helps you recover forgotten instrument passwords; and provides contact methods for technical consultation and after-sales service;
- Zero point calibration, slope calibration and two-point calibration of pH electrode with residual chlorine electrode; measurement error correction function;
- RS485/RS232 communication can be provided according to the user's order statement.



Application area
CLB-H4130 online residual chlorine meter is mainly used in waterworks, drinking water treatment and distribution network, swimming pool, cooling circulating water, water quality monitoring and treatment and other projects to continuously monitor and control the residual chlorine content in aqueous solution.
Hot Tags: Online Residual Chlorine Meter, China Online Residual Chlorine Meter manufacturers, factory, ph ise meter, ph meter probe, ph and chlorine meter, dissolved oxygen content, best professional soil ph meter, iq scientific instruments ph meter