Online Cyanide Analyzer
The DEK-1022 online cyanide analyzer employs a method where tartaric acid and zinc nitrate are added to water samples, followed by heating and distillation. This process allows the distillation of simple cyanides and some complex cyanides into hydrogen cyanide. After absorption by sodium hydroxide, the hydrogen cyanide reacts with cyanamide T to form hydrogen chloride. The reaction then proceeds to a red intermediate product through the reaction with the chromophore isothiocyanic acid-pyrazolone.
- Product Introduction

The DEK-1022 online cyanide analyzer employs a method where tartaric acid and zinc nitrate are added to water samples, followed by heating and distillation. This process allows the distillation of simple cyanides and some complex cyanides into hydrogen cyanide. After absorption by sodium hydroxide, the hydrogen cyanide reacts with cyanamide T to form hydrogen chloride. The reaction then proceeds to a red intermediate product through the reaction with the chromophore isothiocyanic acid-pyrazolone. This red intermediate product is rapidly formed, exhibits maximum absorption at 546nm wavelength, and remains stable for a specified period.
Function description
It has the function of separating clean liquid and waste liquid, and automatically distributes the cleaning liquid and waste liquid to different collection barrels;
- It has automatic calibration function and can be set to execute periodically;
- It has the functions of equal time interval measurement, whole point measurement and controlled measurement to meet the different needs of various fields. It has the automatic standard sample verification function, which can be set to execute on a daily schedule. When the verification is not qualified, it will automatically calibrate and then verify after calibration to ensure the linear accuracy of the equipment;
- It has the automatic cleaning function, and the pipeline and digestion tank will be cleaned before and after the measurement, and can also set regular additional cleaning to ensure the cleanliness of the pipeline and digestion tank;
- It has power-on self-test and abnormal reset functions to ensure that the equipment automatically restores to normal state after power-on after abnormal power failure;
- It has the alarm function of insufficient reagent, heating failure and photoelectric abnormality, which can sound the alarm in real time and record the alarm information;
- It has three levels of permission management function, and different personnel are assigned different operation permissions to prevent non-professionals from misoperation leading to distortion of equipment data or failure;
- It has the function of recording operation logs and process logs, so as to facilitate the view of equipment operation status and personnel operation records.
Technical characteristics
Supports customized range, with multiple options;
- Using multi-point metering module, the quantity is faster and more accurate;
- The new type of discharge valve was jointly developed, which effectively reduced the interference of liquid extraction and improved the service life of the discharge valve; 24-bit high-precision AD chip was used to greatly improve the measurement accuracy;
- The initial signal measurement and measurement can be automatically initialized without manual adjustment, improving the operation and maintenance efficiency;
- The amount of reagent used is greatly reduced, about 40% less than before;
- The data stability is improved, which is relatively increased by 30%-50% compared with the previous;
- System upgrade supports USB drive upgrade, no need for remote, improve operation and maintenance efficiency.
Technical parameter
|
Measuring method: Isonicotinic acid-barbital acid spectrophotometric method |
|||
|
Test range: 0-1mg/L (can be expanded according to customer requirements) |
|||
|
Quantitative lower limit: 0.05 mg/L |
|||
|
Precision : ≤5% |
|||
|
Accuracy: The indication error does not exceed ±10% |
|||
|
Repeatability: Relative standard deviation is not more than 5% |
|||
|
Zero drift: ≤±5% |
|||
|
Range drift: ≤±10% |
|||
|
Environmental temperature stability: ±10% |
|||
|
Voltage stability : ±10% |
|||
|
Memory effect : ±10% |
|||
|
Actual water sample comparison test: ±10% |
|||
|
Communication interface: RS232 digital interface, 4-20mA analog signal output |
|||
|
External dimensions |
485mm*515mm*1260mm |
Weight |
65kg |
|
Power source |
AC(220±10%)V,(50±10%)Hz,5A |
Power |
80W |
|
Ambient temperature |
5~40℃ |
Ambient humidity |
≤90%RH (non-condensation) |
|
Atmospheric pressure |
86 kPa ~ 106 kPa |
Sampling outlets |
Distance L ≤ 15m, drop H ≤ 6m |

Application field
The DEK-1022 online cyanide analyzer is widely used in the field of online monitoring of various pollution sources such as sewage in electronic semiconductor industry, wastewater in electroplating industry, (silver ore) industrial wastewater and surface water.