Treatment Of Abnormal Alarm Of Huihuan Online Total Phosphorus Analyzer
Oct 03, 2025| When the online total phosphorus analyzer is abnormal, it will sound an alarm and interrupt all running programs. After troubleshooting, click the "reset" button, and the instrument will return to normal state.

Common analyzer faults and solutions:
1. Heating anomaly/temperature limit
|
cause |
measure |
|
Temperature setting error |
Change the Settings; |
|
PT100 damaged |
Change the corresponding accessories; |
|
The Line Problem |
Check whether the relevant circuit is poor contact or open circuit; |
2. No reagent/biopsy sample/vineral water was collected, no water sample was collected
|
cause |
measure |
|
Reagents are exhausted |
Make up the corresponding reagent; |
|
Pipeline leak/clogging |
Check the air tightness of the pipeline and check whether there is any blockage in the pipeline; |
|
Peristaltic pump not working |
Check whether the relevant circuit is poor contact or open circuit; Check whether the motor, driver and single chip are damaged; |
|
Electromagnetic valve failure |
Check whether the relevant circuit is poor contact or open circuit; Check whether the solenoid valve, long strip circuit board and single chip microcomputer are damaged; |
3. Photoelectric anomalies
|
cause |
measure |
|
There is sediment or suspended matter in the digestion tank |
Clean the pipeline and add a pre-filter; |
|
Measuring lamp, receiving lamp, photoelectric panel, single chip microcomputer damage |
Change the corresponding accessories; |
|
The Line Problem |
Check whether the relevant circuit is poor contact or open circuit; |
4. Metering anomalies
|
cause |
measure |
|
Measuring white light, measuring black light, photoelectric panel, single chip microcomputer damage |
Change the corresponding accessories; |
|
The Line Problem |
Check whether the relevant circuit is poor contact or circuit break; |
5. The calibration signal cannot be pulled open
|
cause |
measure |
|
Reagent failure, pipeline fluid |
Change reagents and standard samples; check whether the electromagnetic valve of each channel is damaged |
|
Measuring signal problems |
Check whether the measuring lamp, receiving lamp, photoelectric plate and single chip microcomputer are damaged; |
The above abnormal alarm processing is not only suitable for tonline total phosphorus analyzer, but also suitable for volatile phenol analyzel,total zinc analyzer,total lead analyzer,total chromium analyzer,total cadmium analyzer ,total arsenic analyzer and so on.